The new Trek Model PR1182 Electrostatic Voltmeter is a high spatial resolution instrument which is used for a variety of electrostatic applications including contact potential measurements, materials evaluation, and electret studies. The Model PR1182 makes highly accurate, noncontacting measurement of electrostatic voltages of 0 V to ±40 V. The Model PR1182 is specifically designed for high sensitivity voltage measurement applications.
The Model PR1182 has adjustments to compensate for many sensitive testing conditions. A precision calibrated Null Voltage supply of ±10 volts may be used to null different work function (contact potentials) of the test surface or to function as a zero suppression voltage source. A response Speed Control adjusts the speed/noise tradeoff of the Model PR1182 AC response. A self locking Drift/Spacing Null Adjustment minimizes the variation in monitored voltage values as the probe to test surface spacing changes.
An easy to read 3½ digit LED display is provided while measured voltage can also be monitored through a 1:1 voltage monitor output.
The Trek patented low impedance probe assures measurement accuracy which is essentially independent of probe-to-test-surface spacing while eliminating the external environmental effects of high humidity and contamination, such as airborne dust, toner, ions and chemicals, on measurement accuracy.
Measurement Range
0 to ±40 V DC or peak AC.
Spot Spatial Resolution
(at probe-to-surface spacing of 50 um.)
200 um.
Sensitivity
Better than 10 mV.
Accuracy
Voltage Monitor Output
Better than ±0.1% of full scale.
Voltage Display
Better than ±0.1% of full scale,
referred to the voltage monitor.
Speed of Response (10% to 90%)
Less than 4 ms for a 40 V step.
Settling Time
Less than 10 ms.
Stability
Drift with Time
Less than 100 ppm/hour, noncumulative.
Drift with Temperature
Less than 100 ppm/°C.
Response Speed Control
A front panel potentiometer that adjusts the speed of the Model PR1182’s AC response.
Drift/Spacing Null Adjustment
A back panel adjustment minimizes the variation in monitored voltage values as the probe to test surface spacing changes.
Voltage Display
3½ digit LED display.
Range
Switch selectable for ±10 V or ±40 V full scale.
10 V Range Resolution
0.01 V.
40 V Range Resolution
0.1 V.
Zero Offset
±1 count, referred to the voltage monitor.
Sampling Rate
3 readings per second.
Voltage Monitor Output
A buffered 0 to ±40 V output provides a representation
of the measured voltage.
Scale
1:1 of the measured voltage.
Output Noise
Measured using the true rms feature of the Hewlett Packard Model 34401A digital multimeter.
Slow Speed (full counter-clockwise)
Less than 100 mV rms.
Fast Speed (full clockwise)
Less than 200 mV rms.
Output Current
±5 mA (minimum).
Output Impedance
100 ohms, nominal.
Voltage Monitor Output (10:1 ratio)
A buffered 0 to ±10 V output provides a representation of the measured voltage.
Scale
1:10 of the measured voltage.
Output Current
±5 mA (minimum).
Output Impedance
0.1 ohms, nominal
Probe-to-Surface Separation
50 um to 150 um (recommended).
Probe Model PR1182P (round body)
Enclosure Dimensions
9.7 mm diameter x 50 mm (nominal) L
(0.38" diameter x 2" L).
Sensor Orientation
Side.
Cable Length
1524 mm (5 ft).
Weight
3.6 kg (8 lb).
Voltage Monitor Connector
BNC coaxial connector.
Ground Receptacle
Banana jack.
AC Line Cord Receptacle
Standard three-prong line cord with integral fuse holder.
Line Supply
Factory set for one of two voltage ranges:
90 to 127 V AC or 180 to 250 V AC,
at 48 to 63 Hz (specify when ordering).
Operating Conditions
Temperature
0 °C to 40 °C.
Relative Humidity
To 90%, noncondensing.
Power Switch
A two position power ON/OFF switch.
TREK, INC. certifies that each Model PR1182 is tested and calibrated to specifications using measurement equipment traceable to the National Institute of Standards and Technology or traceable to consensus standards.
Low Voltage Safety Compliance
(EN 61010-1)
Overvoltage Category
CAT II: Local-level mains, appliances, portable equipment.
Pollution Category
Degree 1: Operate in environments where no pollution or only dry nonconductive pollution occurs.
All specifications are subject to change.