High Voltage Power Supply


TREK Model 610E High Voltage Power Supply Amplifier Controller

The Trek Model 610E is a high voltage power supply, amplifier, and controller that is used in many applications including closed-loop charge control, electrophotographic research, insulation testing, dielectric material evaluation, and as an amplifier for AC or DC calibrators and power supplies. The Model 610E has an output voltage range of 0 V to ±10 kV with an output current range of 0 to ±2000 µ A, and can be operated in any of the six modes.

As a high-voltage amplifier, the Model 610E amplifies an externally applied signal. The gain is switch selectable between 100 V/V or 1000 V/V. As a high-voltage reference power supply, a front panel dial commands the output voltage.

As a transconductance amplifier, an externally applied voltage signal produces a proportional output current. The transconductance gain is switch selectable between 20 µ A/V or 200 µ A/V.

As a current supply, a front panel dial commands the output current. As a high-voltage controller, the high-voltage amplifier mode is maintained, but the amplifier input and feedback elements are uncommitted and available for configuration by the user.

As a transconductance controller, the transconductance amplifier mode is maintained, but the amplifier input and feedback elements are uncommitted and available for configuration by the user.

The Model 610E features a compliance limit function. When in a CURRENT command mode, the compliance function is voltage. When in a VOLTAGE command mode, the compliance function is current. The current compliance limit function can be based upon either the total current or a specific load current path.

The Model 610E features a four-quadrant, active output stage that sinks or sources current into reactive or resistive loads throughout the output voltage range. This is essential for achieving the accurate output response and high slew rates demanded by reactive loads. The Model 610E is protected against overvoltage and overcurrent conditions that may be generated by active loads or by output short circuits to ground.

Precision voltage and current monitors provide low-voltage representations of the high-voltage output and load current for monitoring purposes, or for use as feedback signals in a closed-loop system. The current control loop provide a means for close regulation of either the total load current, I(t), or a specific current path, I(r). When the load includes several current paths, one may be chosen, I(r), to be closely regulated.

Show/Hide Product Technical Information

PRODUCT APPLICATIONS

  • USED AS A TRANSCONDUCTANCE AMPLIFIER AND CONTROLLER
    • Output Current Range
      0 to ±200 µA or 0 to ±2000 µA DC or peak AC, switch selectable.
    • Input Voltage Range
      0 to ±10 V DC or peak AC.
    • Transconductance Gain 200 µA range
      20 µA/V.
    • Transconductance Gain 2000 µA range
      200 µA/V.
    • DC Current Gain Accuracy
      Better than 0.3% of full scale, max (Controller mode Gain Accuracy is dependent on user-specified components.)
    • Compliance Voltage Range
      0 to ±10 kV DC or peak AC (adjustable using a precision potentiometer with a dial).
    • Offset Current
      Less than 0.1 µA.
  • USED AS A HIGH-VOLTAGE POWER SUPPLY
    • Output Voltage Range
      0 to +1 kV, 0 to -1 kV, 0 to +10 kV, or 0 to -10 kV, switch selectable and adjustable using a precision potentiometer with a dial.
    • Resolution 1 kV range
      1 volt.
    • Resolution 10 kV range
      10 volts.
    • Accuracy
      Better than 0.1% of full scale, typical.
    • Compliance Current Range (adjustable)
      0 to ±2 mA DC, using a precision potentiometer with a dial.
    • Output Noise
      Less than 700 mV rms (measured with a
      20 kHz true rms meter).
    • Regulation
    • Line
      Better than 0.01% for all line voltages.
    • Load
      Better than 0.01% for a load change from
      1 µA to 1000 µA.
    • Stability
    • Drift with Temperature
      Less than 100 ppm/ °C.
    • Drift with Time
      Less than 100 ppm /24 hours.
  • USED AS A CURRENT SUPPLY
    • Output Current Range
      0 to 200 µA, or 0 to 2000 µA, switch selectable and adjustable using a precision potentiometer with a dial.
    • Resolution 200 µA range
      0.2 µA.
    • Resolution 2000 µA range
      2 µA.
    • Accuracy
      Better than 0.3% of full scale, typical.
      Better than 0.5% of full scale, max.
    • Compliance Voltage Range (adjustable)
      0 to ±10 kV DC, adjustable using a precision potentiometer with a dial.
    • Regulation
    • Line
      Better than 0.01% for all line voltages.
    • Load
      Better than 0.01% for a load change from10 V to 10 kV.
    • Stability
    • Drift with Temperature
      Less than 100 ppm/°C.
    • Drift with Time
      Less than 100 ppm/24 hours.
  • USED AS A HIGH-VOLTAGE AMPLIFIER AND CONTROLLER
    • Output Voltage Range
      0 to ±1 kV or 0 to ±10 kV DC or peak AC, switch selectable.
    • Input Voltage Range
      0 to ±10 V DC or peak AC.
    • Voltage Gain 1 kV Range
      100 V/V.
    • Voltage Gain 10 kV Range
      1000 V/V.
    • DC Voltage Gain Accuracy
      Better than 0.3% of full scale (Controller mode Gain Accuracy is dependent on user-specified components).
    • Compliance Current Range
      0 to ±2 mA DC or peak AC (adjustable using a precision potentiometer with a dial).
    • Offset Voltage
      Less than 2 V.
    • Output Noise
      Less than 700 mV rms (measured with a 20 kHz true rms meter.
    • Slew Rate (10% to 90%, typical)
      Greater than 20 V/µs.
    • Large Signal Bandwidth (-3dB)
      DC to greater than 1.2 kHz.
    • Large Signal Bandwidth (1% Distortion)
      DC to greater than 600 Hz.
    • Small Signal Bandwidth (-3dB)
      DC to 10 kHz.
    • Settling Time to 1%
      Less than 1 ms for a 10 kV step.
    • Input Impedance
    • Noninverting Configuration
      20 kohm, nominal.
    • Inverting Configuration
      10 kohm, nominal.
    • Differential Configuration
      10 kohm, nominal.

PRODUCT FEATURES

    • Amplifier Input Configuration Programming
      The 610E high-voltage power supply can be configured as a noninverting amplifier, an inverting amplifier, or a differential amplifier.
    • External Control Connector
      The high-voltage supply can be turned ON or OFF remotely using a multi-pin connector on the rear of the unit.
    • Digital Enable
      An input providing a connection for a TTL compatible signal to drive the high-voltage power supply output toward zero or allow normal high-voltage operation.
    • Compliance Feature
      • Compliance Level Selection
        A precision potentiometer with a calibrated dial used to set the current limit when operating in the voltage mode, or sets a voltage limit when operating in the current mode.
      • Compliance Indicator
        A yellow LED will illuminate when the instrument is in a compliance limit condition.
      • Compliance Limit
      • Current mode
        Adjustable to within 20 V of the output setting.
      • Voltage mode
        Adjustable to within 0.5 µA of the output setting.
    • High-Voltage Switch
      A three-position switch to select ON, OFF or REMOTE.
    • Mode Control Switch
      A three position switch to select either the High-Voltage Power Supply, Amplifier, or Controller mode of operation
    • Supply Mode Voltage Control
      • Range Selection
        A two-position switch to select 0 to ±1 kV or 0 to ±10 kV.
      • Output Level Selection
        A precision potentiometer with a graduated dial.
      • Polarity Selection
        A three-position switch to select positive (+), negative (-), or OFF.
    • Front Panel Display
      A 3 1/2-digit LED display can be selected to indicate the voltage output or the current output.
      • Voltage Range
        0 to ±1 kV or 0 to ±10 kV (switch selectable).
      • Current Range
        0 to ±199.9 µA or 0 to ±1999 µA (switch selectable).
      • Voltage Resolution
      • 1 kV Range
        1 V.
      • 10 kV Range
        10 V.
      • Current Resolution
      • 200 µA Range
        0.1 µA.
      • 2000 µA Range
        1 µA.
      • Zero Offset
        ±1 count, referred to the Voltage and Current monitors.
    • Voltage Monitor
      • Scale
        1/1000th of the output voltage
      • DC Gain Accuracy
        Better than 0.1% of full scale as referred to the high-voltage output.
      • Offset Voltage
        Less than 5 mV.
      • Noise
        Less than 20 mV p-p.
      • Output Impedance
        47 ohms, nominal.
    • Current Monitor
      • Scale
        1 V/200 µA
      • Noise
        Less than 30 mV p-p.
      • DC Gain Accuracy
        Better than 0.1% of full scale as referred to the high-voltage output.
      • Offset Voltage
        Less than 10 mV.
      • Output Impedance
        1 kohm, nominal.

Certification


    TREK, INC. certifies that each Model 610E is tested and calibrated to specifications using measurement equipment traceable to the National Institute of Standards and Technology or traceable to consensus standards. A Certificate of Calibration accompanies each instrument when it is shipped from the factory.

    Low Voltage Safety Compliance (EN61010-1)

    Installation Category
    CAT II: Local-level mains, appliances, portable equipment.

    Pollution Category
    Degree 1: Operate in environments where no pollution occurs or only dry, nonconductive pollution occurs.All specifications are with no load unless otherwise noted.

    All Specifications are with no load unless otherwise noted.
    All specifications are subject to change.