Model 601C DC-Stable High-Voltage Power Amplifier
Now equipped with SHV type high-voltage output connectors.
The Trek Model 601C is a DC-stable, high-voltage power amplifier designed to provide precise control of output voltages in the ranges of:
0 to ±500 V, DC or peak AC, 0 to -1 kV, DC or peak AC,
0 to +1kV, DC or peak AC.
With an output current in the range of: 0 to ±10 mA DC, 0 to ±20 mA peak AC.
The Model 601C features an all solid-state design for high slew rate, wide bandwidth, and low-noise operation. The four-quadrant, active output stage sinks or sources current into reactive or resistive loads throughout the output voltage range. This is essential for achieving the accurate output response and high slew rates demanded by reactive loads.
The Model 601C is protected against overvoltage and overcurrent conditions that may be generated by active loads or by output short circuits to ground.
Precision voltage and current monitors provide low-voltage representations of the high-voltage output and load current for monitoring purposes or for use as feedback signals in a closed-loop system.
The Digital Enable feature provides a connection for a remote device to turn ON and OFF the high voltage of the instrument. This makes the Model 601C ideal for automated or computer controlled systems. The Model 601C has a Dynamic Adjustment feature which can be used to optimize the AC response of the output signal.
The Model 601C is available in single channel or dual channels configurations. These units can be operated on a bench top or, with optional hardware, in a standard 19-inch rack.
The Model 601C can be operated as a noninverting or inverting amplifier with a fixed gain of 100 V/V. The input gain of the 601C can be programmed for 50 V/V for the ±500 V range only. The Model 601C comes from the factory with settings for an output voltage of ±500 V DC or peak AC, an input signal ratio of 100:1 V/V, and a noninverting input.
back to top
PERFORMANCE
- DC Voltage Gain
100 V/V (a gain of 50 V/V is available for the ±500 V range only). Specify when ordering or contact TREK, INC. for more information.
- DC Voltage Gain Accuracy
Better than 0.1% of full scale.
- Offset Voltage
Less than 500 mV.
- Output Noise
Less than 10 mV rms (measured using the true rms feature of the Hewlett Packard Model 34401A digital multimeter).
- Slew Rate (10% to 90%, typical)
Greater than 50 V/µs.
- Large Signal Bandwidth (1% distortion)
DC to 8 kHz.
- Small Signal Bandwidth (-3dB)
DC to 30 kHz.
- Settling Time to 1%
Less than 300 µs for a 1 kV step.
- Stability
- Drift with Time
Less than 100 ppm/hr, noncumulative.
- Drift with Temperature
Less than 50 ppm/°C.
back to top
PRODUCT FEATURES
- If more than one output voltage range, input configuration, or input gain ratio is required, please contact TREK, INC. for more information.
- The 601C can have up to two (2) independent amplifier channels in one enclosure. The following features are specific to each amplifier channel.
- Output Voltage Configurations
The high-voltage output can be programmed for
0 to ±500 V, 0 to -1 kV, or 0 to +1 kV DC or peak AC.
Specify when ordering or contact
TREK, INC. for more information.
- Input Configuration
The 601C is factory configured as a noninverting amplifier. An inverting amplifier configuration is available. Specify when ordering or contact TREK INC. for more information
- Dynamic Adjustment
A graduated one-turn potentiometer is used to optimize the AC response of the Model 601C under various load configurations.
- Digital Enable
An input providing a connection for a TTL compatible signal to turn on and off the high-voltage output. A TTL high (or open) turns off the high-voltage output. A TTL low turns on the high-voltage output.
- Voltage Monitor
A buffered output providing a low-voltage replica of the high-voltage output
- Scale
1/100th of the high-voltage output.
- DC Accuracy
Better than 0.1% of full scale.
- Offset Voltage
Less than 5 mV.
- Output Noise
Less than 10 mV rms (measured using the true rms feature of the Hewlett Packard Model 34401A digital multimeter).
- Output Impedance
Less than 0.1 ohm.
- Current Monitor
A buffered output providing a low-voltage representation of the load current.
- Scale
0.5 V/mA.
- DC Accuracy
Better than 1% of full scale.
- Offset Voltage
Less than 10 mV.
- Output Noise
Less than 20 mV (measured using the true rms feature of the Hewlett Packard Model 34401A digital multimeter).
- Output Impedance
Less than 0.1 ohm.
back to top
OUTPUT
- Output Voltage Range
Factory set for 0 to ±500 V.
- Available output voltage ranges are
0 to -1 kV or 0 to +1 kV DC or peak AC.
Specify when ordering or contact
TREK, INC. for more information.
- Output Current Range
0 to ±10 mA DC.
0 to ±20 mA peak AC.
back to top
AMPLIFIER INPUT
- Input Voltage Range
0 to ±10 V DC or peak AC.
- Input Impedance
25 k ohm, nominal
back to top
ACCESSORIES
- Accessories Supplied
- Operator’s manual, SHV mating connector for each amplifier channel, line cord.
back to top
GENERAL
- Dimensions
- Single Channel Instrument
width - 220.5 mm (8.68 in.)
height - 88.4 mm (3.48 in.)
depth - 335 mm (13.19 in.)
- Dual Channel Instrument
width - 433.8 mm (17.08 in.)
height - 88.4 mm (3.48 in.)
depth - 335 mm (13.19 in.)
- Weight
- Single Channel Instrument
4.3 kg (9.4 lb)
- Dual Channel Instrument
8.6 kg (18.8 lb)
- High-Voltage Output Connector
SHV high-voltage connector.
- Amplifier Input
BNC connector.
- Voltage Monitor
BNC connector.
- Current Monitor
BNC connector.
- Digital Enable Connector
BNC connector.
- Power Receptacle
Standard three-prong power connector with an integral fuse holder.
- Power Requirements
- Line Voltage
Factory set for one of two ranges: 90 to 127 V AC or 180 to 250 V AC, at 50/60 Hz. (Specify when ordering.)
- Power Consumption
150 VA, maximum.
- Operating Conditions
- Temperature
0 °C to 40 °C.
- Relative Humidity
To 85%, noncondensing.
back to top
CERTIFICATION
TREK, INC. certifies that each Model 601C is tested and calibrated to specifications using measurement equipment traceable to the National Institute of Standards and Technology or traceable to consensus standards. A Certificate of Calibration accompanies each instrument when it is shipped from the factory.
All specifications are with no load unless otherwise noted.
All specifications are subject to change.
back to top